Instrument QC
Overview of Instrument Quality Control (QC)
- Definition: Instrument quality control (IQC) refers to the procedures and practices used to ensure that a flow cytometer is operating according to its specifications and is producing accurate and reliable data
-
Importance:
- Data Accuracy: To ensure that the data accurately reflects the biological system being studied
- Reproducibility: To ensure that experiments can be reproduced over time
- Compliance: To comply with regulatory requirements and accreditation standards
-
Key Aspects of Instrument Quality Control:
- Optical Alignment
- Detector Calibration
- Fluidics System Checks
- Electronic Stability
Optical Alignment
- Definition: The process of ensuring that the lasers, lenses, mirrors, and detectors in the flow cytometer are properly aligned
-
Importance:
- Optimal Signal: To maximize the amount of light that reaches the detectors
- Consistent Performance: To ensure consistent performance of the flow cytometer over time
- Accurate Measurements: To ensure that the fluorescence and scatter measurements are accurate
-
Methods:
- Visual Inspection:
- Inspect the optical components for any signs of damage or misalignment
- Look for any obstructions or debris in the light path
- Alignment Beads:
- Use alignment beads to optimize the position of the lasers and detectors
- Adjust the alignment until the beads are centered in the flow cell and the signal intensity is maximized
- Software Tools:
- Use software tools to monitor the alignment of the optical system over time
- The software may provide guidance on how to adjust the alignment
- Visual Inspection:
-
Troubleshooting Common Optical Misalignment Problems:
- Weak Signals:
- Make sure there are no obstructions, and verify laser function.
- Variability:
- Use the QC beads to assess instrument performance.
- Clean, repair, or replace parts, as needed.
- Weak Signals:
-
General notes:
- Optical alignment should occur when there are any known issues or at a fixed interval (e.g. weekly)
Detector Calibration
- Definition: The process of adjusting the voltages of the detectors (PMTs or APDs) to ensure that they are responding linearly to changes in light intensity
-
Importance:
- Accurate Quantification: To ensure that the fluorescence values accurately reflect the expression of the target antigens
- Data Comparison: To allow for comparison of data from different experiments or different instruments
-
Methods:
-
Calibration Beads:
- Use a set of calibration beads with known fluorescence intensities
- The calibration beads are typically labeled with a range of fluorescence intensities that span the range of the flow cytometer
-
Voltage Adjustment:
- Run the calibration beads on the flow cytometer
- Adjust the detector voltages until the fluorescence intensity values of the beads match their known values
-
Linearity Check:
- Verify that the detectors are responding linearly to changes in light intensity
- This can be done by plotting the fluorescence intensity values of the calibration beads against their known MESF values
-
Calibration Beads:
- Linearity of Voltage should be carefully checked
Fluidics System Checks
- Definition: The process of verifying that the fluidics system is functioning properly
-
Importance:
- Stable Flow Rate: To ensure a stable and consistent flow rate, which is essential for accurate cell counting and sizing
- Proper Focusing: To ensure that the cells are properly focused, which is essential for optimal signal intensity and resolution
- Prevention of Clogging: To prevent clogging of the flow cell or tubing, which can disrupt the flow of cells
-
Methods:
- Flow Rate Measurement:
- Measure the flow rate using a graduated cylinder and a timer
- Compare the measured flow rate to the expected flow rate
- Bubble Check:
- Inspect the tubing and flow cell for air bubbles
- Remove any air bubbles that are present
- Pressure Check:
- Monitor the pressure readings in the fluidics system
- Ensure that the pressure is within the acceptable range
- Flow Rate Measurement:
-
Examples of tests for fluidics system checks:
- Verify that the sheath fluid is at the appropriate level
- Remove any air bubbles
- Clean the instrument with detergent, especially after running sticky samples
Electronic Stability
- Definition: Assessing the stability of the flow cytometer’s electronics, which can affect the accuracy and reproducibility of the measurements
-
Importance:
- Ensure that the data is consistent.
- To allow results to be properly interpreted and to troubleshoot other aspects.
-
Electronic stability can be affected by:
- Room temperature fluctuation
- Variances from normal conditions
-
How to troubleshoot:
- Assess the flow rate and stability with calibration beads before and after running samples
Troubleshooting Instrument Quality Control Issues
-
Weak Signals:
-
Possible Causes:
- Misaligned optics
- Low laser power
- Incorrect detector settings
-
Troubleshooting Steps:
- Align the optical system
- Check laser power
- Adjust detector settings
-
Possible Causes:
-
High Background Noise:
-
Possible Causes:
- Contaminated reagents
- Incorrect instrument settings
- Electrical interference
-
Troubleshooting Steps:
- Use clean reagents
- Optimize instrument settings
- Reduce electrical interference
-
Possible Causes:
-
Variable Results:
-
Possible Causes:
- Inconsistent technique
- Instrument instability
- Sample heterogeneity
-
Troubleshooting Steps:
- Standardize IQC protocols
- Calibrate instrument
- Increase sample size
-
Possible Causes:
Key Terms
- Instrument Quality Control (IQC): The process of ensuring that a flow cytometer is operating according to its specifications
- Optical Alignment: Ensuring that the lasers, lenses, mirrors, and detectors are properly aligned
- Detector Calibration: Adjusting the voltages of the detectors to ensure that they are responding linearly to changes in light intensity
- Fluidics System Checks: Verifying that the fluidics system is functioning properly
- Electronic Stability: Assessing the stability of the flow cytometer’s electronics
- Calibration Beads: Beads with known fluorescence intensities that are used to calibrate the flow cytometer