Instrument QC

Overview of Instrument Quality Control (QC)

  • Definition: Instrument quality control (IQC) refers to the procedures and practices used to ensure that a flow cytometer is operating according to its specifications and is producing accurate and reliable data
  • Importance:
    • Data Accuracy: To ensure that the data accurately reflects the biological system being studied
    • Reproducibility: To ensure that experiments can be reproduced over time
    • Compliance: To comply with regulatory requirements and accreditation standards
  • Key Aspects of Instrument Quality Control:
    • Optical Alignment
    • Detector Calibration
    • Fluidics System Checks
    • Electronic Stability

Optical Alignment

  • Definition: The process of ensuring that the lasers, lenses, mirrors, and detectors in the flow cytometer are properly aligned
  • Importance:
    • Optimal Signal: To maximize the amount of light that reaches the detectors
    • Consistent Performance: To ensure consistent performance of the flow cytometer over time
    • Accurate Measurements: To ensure that the fluorescence and scatter measurements are accurate
  • Methods:
    • Visual Inspection:
      • Inspect the optical components for any signs of damage or misalignment
      • Look for any obstructions or debris in the light path
    • Alignment Beads:
      • Use alignment beads to optimize the position of the lasers and detectors
      • Adjust the alignment until the beads are centered in the flow cell and the signal intensity is maximized
    • Software Tools:
      • Use software tools to monitor the alignment of the optical system over time
      • The software may provide guidance on how to adjust the alignment
  • Troubleshooting Common Optical Misalignment Problems:
    • Weak Signals:
      • Make sure there are no obstructions, and verify laser function.
    • Variability:
      • Use the QC beads to assess instrument performance.
      • Clean, repair, or replace parts, as needed.
  • General notes:
    • Optical alignment should occur when there are any known issues or at a fixed interval (e.g. weekly)

Detector Calibration

  • Definition: The process of adjusting the voltages of the detectors (PMTs or APDs) to ensure that they are responding linearly to changes in light intensity
  • Importance:
    • Accurate Quantification: To ensure that the fluorescence values accurately reflect the expression of the target antigens
    • Data Comparison: To allow for comparison of data from different experiments or different instruments
  • Methods:
    1. Calibration Beads:
      • Use a set of calibration beads with known fluorescence intensities
      • The calibration beads are typically labeled with a range of fluorescence intensities that span the range of the flow cytometer
    2. Voltage Adjustment:
      • Run the calibration beads on the flow cytometer
      • Adjust the detector voltages until the fluorescence intensity values of the beads match their known values
    3. Linearity Check:
      • Verify that the detectors are responding linearly to changes in light intensity
      • This can be done by plotting the fluorescence intensity values of the calibration beads against their known MESF values
  • Linearity of Voltage should be carefully checked

Fluidics System Checks

  • Definition: The process of verifying that the fluidics system is functioning properly
  • Importance:
    • Stable Flow Rate: To ensure a stable and consistent flow rate, which is essential for accurate cell counting and sizing
    • Proper Focusing: To ensure that the cells are properly focused, which is essential for optimal signal intensity and resolution
    • Prevention of Clogging: To prevent clogging of the flow cell or tubing, which can disrupt the flow of cells
  • Methods:
    • Flow Rate Measurement:
      • Measure the flow rate using a graduated cylinder and a timer
      • Compare the measured flow rate to the expected flow rate
    • Bubble Check:
      • Inspect the tubing and flow cell for air bubbles
      • Remove any air bubbles that are present
    • Pressure Check:
      • Monitor the pressure readings in the fluidics system
      • Ensure that the pressure is within the acceptable range
  • Examples of tests for fluidics system checks:
    • Verify that the sheath fluid is at the appropriate level
    • Remove any air bubbles
    • Clean the instrument with detergent, especially after running sticky samples

Electronic Stability

  • Definition: Assessing the stability of the flow cytometer’s electronics, which can affect the accuracy and reproducibility of the measurements
  • Importance:
    • Ensure that the data is consistent.
    • To allow results to be properly interpreted and to troubleshoot other aspects.
  • Electronic stability can be affected by:
    • Room temperature fluctuation
    • Variances from normal conditions
  • How to troubleshoot:
    • Assess the flow rate and stability with calibration beads before and after running samples

Troubleshooting Instrument Quality Control Issues

  • Weak Signals:
    • Possible Causes:
      • Misaligned optics
      • Low laser power
      • Incorrect detector settings
    • Troubleshooting Steps:
      • Align the optical system
      • Check laser power
      • Adjust detector settings
  • High Background Noise:
    • Possible Causes:
      • Contaminated reagents
      • Incorrect instrument settings
      • Electrical interference
    • Troubleshooting Steps:
      • Use clean reagents
      • Optimize instrument settings
      • Reduce electrical interference
  • Variable Results:
    • Possible Causes:
      • Inconsistent technique
      • Instrument instability
      • Sample heterogeneity
    • Troubleshooting Steps:
      • Standardize IQC protocols
      • Calibrate instrument
      • Increase sample size

Key Terms

  • Instrument Quality Control (IQC): The process of ensuring that a flow cytometer is operating according to its specifications
  • Optical Alignment: Ensuring that the lasers, lenses, mirrors, and detectors are properly aligned
  • Detector Calibration: Adjusting the voltages of the detectors to ensure that they are responding linearly to changes in light intensity
  • Fluidics System Checks: Verifying that the fluidics system is functioning properly
  • Electronic Stability: Assessing the stability of the flow cytometer’s electronics
  • Calibration Beads: Beads with known fluorescence intensities that are used to calibrate the flow cytometer